The ScanRunner consume loop runs on the same thread that drains the live
USB histogram batch queue, so a slow disk/DB sink (SQLite commit/fsync, CSV
flush) stalls that drain, backpressures the parser and dev.read, and
overflows the sensor firmware's 4-deep histo queue — dropping histogram
frames. Root-caused on the 2026-06 long-scan soak ("HISTO enqueue fail:
queue full ... enq-deq=4 txfail=0": brief host-side drain stalls, not a USB
fault and not the persistent histo-wedge).
Fix #1 — decouple storage sinks from the drain:
- New AsyncSink wraps a sink and runs consume() on a dedicated worker
thread fed by a bounded FIFO buffer, so a transient sink stall is
absorbed instead of stalling the USB path. on_scan_start stays
synchronous (preserves critical-sink abort + ordering); on_complete
drains the buffer and joins before finalising, so no persisted data is
lost. ScanWorkflow now wraps CsvSink and ScanDBSink.
- LiveUsbSource._batch_queue depth is configurable and defaulted up from 4
(~1 s) to 16 so other runner hiccups (pipeline CPU spike, GC) are
absorbed before reaching dev.read.
Fix #3 — make sub-second host backpressure visible:
- StreamInterface times every packet-queue put. A put that blocks >=100 ms
(previously only the >1 s drop was logged) is counted per scan
(backpressure_stalls / backpressure_stall_s / dropped_chunks),
rate-limit warned, and summarised at stop_streaming.
(Firmware queue-depth bump is intentionally out of scope — RAM-bound.)
Tests: AsyncSink (FIFO ordering, drain-before-finalise, critical-abort
propagation, worker error isolation, Sink protocol, runner transparency,
wrapped accessor), configurable batch-queue depth, StreamInterface
backpressure counters.
Co-Authored-By: Claude Opus 4.8 <noreply@anthropic.com>
Replace the magic batch-queue default (16) with a value derived from a 3 s buffer target and the capture rate (CAPTURE_HZ) + batch size, so the intent is self-documenting and survives a different batch_size_frames. At the default 10-frame batch this is 12 batches (~3 s); the 64-deep packet queue adds ~1.6 s behind it. batch_queue_size stays overridable. Co-Authored-By: Claude Opus 4.8 <noreply@anthropic.com>
Contributor
Author
boringethan
added a commit
that referenced
this pull request
Sep 29, 2026
…covery Brings the merged OB-telemetry work (PR #170 / sensor-fw #104, wire version 2) into the bench branch so the parser matches the flashed firmware 1.8.2-dev.1-11-g93f3ca4 — the version/size guard rejects every v1 snapshot otherwise. Adds z_avg (the four zero-line dark-row averages, /64 -> DN), z_avg_mean/spread, blk_lvl_target, blc_thres, the black/zero-line windows, and the BLC fault latches to get_camera_telemetry() and CameraTelemetryCsvLogger. Also pulls the HISTO-pipe recovery (PR #172 / sensor-fw #106) so a telemetry poll that stalls a reader no longer wedges the histogram stream (issue #96). Files taken verbatim from origin/next (my branch never diverged them). Bench validation on the RIGHT sensor: z_avg_mean = 126.6 DN pinned to the 128 servo target, spread 0.5 DN, fault latch clean; histograms + telemetry both survived a 60 s co-polled scan. 15/15 telemetry unit tests pass. Co-Authored-By: Claude Opus 4.8 <noreply@anthropic.com>
boringethan
added a commit
that referenced
this pull request
Sep 29, 2026
The first warm-restart run (WRESTART) failed after the cold reference scan: every powered-idle handoff timed out on connect (Errno 10060). Root cause is the sensor-fw #96 COMM stall -- sustained 1 Hz camera-telemetry polling wedges the command endpoint ~9-18 min into a scan. A rig-OFF cooldown hides it because the next scan's power cycle recovers the endpoint; a powered-idle gap keeps the rig ON, so the stall carries into the next connect and it never recovers. The raw histogram stream (die temp + intensity) is unaffected and is all the warm-restart ladder needs, so telemetry is now auto-disabled with a warning whenever any ladder step is powered-idle. Also: powered-idle now follows the step's fan_on flag instead of forcing the fan off, so the gap can cool the module from the streaming plateau at a chosen rate (fan-on gives an informative spread of restart dip depths; fan-off barely cools). Co-Authored-By: Claude Opus 4.8 <noreply@anthropic.com>
boringethan
added a commit
that referenced
this pull request
Oct 1, 2026
1 Hz camera-telemetry polling alongside the image stream stalled the sensor COMM endpoint ~3 min into FFDIP_01 (USBTimeoutError 10060, cf. sensor-fw #96), which also blinds thermal_soak's --max-die-c stop. Use the soak's 10 s default. Co-Authored-By: Claude Opus 5.5 <noreply@anthropic.com>
This file contains hidden or bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
Add this suggestion to a batch that can be applied as a single commit.This suggestion is invalid because no changes were made to the code.Suggestions cannot be applied while the pull request is closed.Suggestions cannot be applied while viewing a subset of changes.Only one suggestion per line can be applied in a batch.Add this suggestion to a batch that can be applied as a single commit.Applying suggestions on deleted lines is not supported.You must change the existing code in this line in order to create a valid suggestion.Outdated suggestions cannot be applied.This suggestion has been applied or marked resolved.Suggestions cannot be applied from pending reviews.Suggestions cannot be applied on multi-line comments.Suggestions cannot be applied while the pull request is queued to merge.Suggestion cannot be applied right now. Please check back later.
Refs #116. Superseded by #113 (same work rebased + review fixes); kept open as fallback until #113 merges, then this closes.
Why
On the 2026-06 long-scan soak the sensor firmware logged ~833
HISTO enqueue fail: queue fullevents (enq-deq=4,txfail=0). Root cause: theScanRunnerconsume loop runs on the same thread that drains the live USB histogram batch queue, so a slow disk/DB sink — a SQLite commit/fsync or a CSV flush — stalls that drain. The stall backpressures the parser →dev.read→ the firmware's deliberately shallow 4-deep histo queue (~100 ms at 40 Hz) overflows and frames are dropped. These were brief (>100 ms, <1 s) host-side stalls, not a USB fault and not the persistent histo-wedge.What
Fix #1 — decouple the storage sinks from the USB drain
AsyncSink(omotion/pipeline/async_sink.py) wraps a sink and runsconsume()on a dedicated worker thread fed by a bounded FIFO buffer, so a transient sink stall is absorbed by the buffer instead of stalling the USB path.on_scan_startstays synchronous (preserves critical-sink abort + ordering before the first consume);on_completedrains the buffer and joins the worker before finalising the wrapped sink, so no persisted data is lost. Sustained overload still backpressures once the buffer fills (correct — you can't outrun the disk forever).ScanWorkflownow wrapsCsvSinkandScanDBSinkinAsyncSink(the live-plot/diagnostics sinks stay synchronous so the GUI trace isn't delayed).LiveUsbSource._batch_queuedepth is now configurable and defaulted up from 4 (~1 s) to 16, so other runner hiccups (pipeline CPU spike, GC) are absorbed before reachingdev.read.Fix #3 — make the previously-invisible sub-second backpressure observable
StreamInterfacenow times every packet-queue put. A put that blocks ≥100 ms (previously only the >1 s drop was logged) is counted per scan —backpressure_stalls/backpressure_stall_s/dropped_chunks— rate-limit warned mid-scan, and summarised in thestop_streamingline (logged at WARNING if any host-side drops occurred).Out of scope: bumping the firmware
HISTO_QUEUE_SIZE(4→8) — RAM-bound on the sensor, intentionally left alone. This PR removes the host-side cause instead.Tests
New, test-first:
test_async_sink.py— FIFO ordering, drain-before-finalise, critical-abort propagation, worker error isolation,Sinkprotocol conformance, transparency through a realScanRunner,wrappedaccessor.test_stream_interface_backpressure.py— no-stall fast path, sub-second stall counting, drop-past-timeout, counter reset onstart_streaming.test_sources.py— configurable / deepened_batch_queue.Full pipeline + scan-workflow suites pass. (The 3 pre-existing
ScanDBSink"empty session deleted" failures onnextare unrelated to this change and untouched here.)🤖 Generated with Claude Code