SEEDLESS frames: first-N-frames unseeded-TA scan mode (host-side) #146

Open

Description

@boringethan

SDK side of OpenwaterHealth/openmotion-bloodflow-app#361 (SEEDLESS Frames, formerly NO_TA). Experimental engineering-test feature requested by @bahartl: run the first N frames of a scan with the seed off, the TA pulsing 2 ms (deliberately inducing TA self-lasing), and camera exposure at 2295 µs, then revert to normal scan parameters mid-scan.

Approved design spec: https://github.com/OpenwaterHealth/openmotion-sdk/blob/claude/github-issue-link-3ca73e/docs/superpowers/specs/2026-07-15-seedless-frames-design.md (laser-engineer sign-off 2026-07-16).

Everything is host-side register writes — no firmware or FPGA changes (the TA FPGA is an edge-triggered one-shot, so TA_PULSE_WIDTH is writable mid-scan without a trigger restart).

Scope:

  • ScanRequest.seedless_frames: int = 0 (0 = off)
  • omotion/seedless.py — SeedlessController: register snapshot via I2C read-back, pre-trigger apply (seed gains → 0, TA → raw 6250, EE/OPT_PULSE_WIDTH_UL widened to raw 7813, exposure 0xFF), frame-ID-driven restore at frame N with strict ordering (TA width → 50 ms wait → safety limits → seed gains → exposure; reverse order latches TA_shutdown), idempotent force-restore on all teardown/cancel paths
  • ScanWorkflow hooks: pre-trigger apply, on_row frame-N watch, finally/cancel restore; abort pre-trigger if apply fails
  • Trigger config: LaserPulseSkipDelayUsec → 2500 when seedless, so scheduled dark frames stay outside the 2295 µs exposure window
  • FrameClassificationStage: tag frames 1..N seedless, N+1..N+80 seedless_tx (guard band for exposure smear); dtype <U8 → <U12; excluded from BFI/BVI + dark correction like warmup; N recorded in scan metadata
  • Unit tests (no hardware): restore ordering, latch idempotency, snapshot-fallback, classification tags
  • Bench verification per spec §Testing

Deliberate safety note: the 2 ms TA pulse exceeds the 1.000 ms safety-FPGA pulse-width upper limit; widening the UL for the seedless window was explicitly authorized by Ethan for this engineering test. The controller restores baseline limits on every exit path, and a console power-cycle also clears them (volatile registers).

App-side UI (engineering-mode checkbox + N spinner) stays on OpenwaterHealth/openmotion-bloodflow-app#361.

Activity

  1. 19 remaining items

  2. boringethan commented on Jul 18, 2026

    @boringethan
    ContributorAuthor

    Bench session 2026-07-17 (evening): DEBUG_FLAG_CAMERA_CROP × seedless, rebase, two structural fixes

    Branch rebased onto latest next (clean; suite grew to 573 passing) and the seedless experiment re-run with the camera crop (1720×1280, right 200 columns dropped) per Ethan's request. All four regime × crop combinations captured on sensor-fw 1.8.2-dev.1, N=25, mask 0x66.

    Crop effect on the contrast numbers (frames 11..25 seedless vs light, pedestal-subtracted)

    scan (left / right) seedless K light K separation (light K / seedless K)
    boosted, full frame 0.095 / 0.089 0.399 / 0.406 4.19 / 4.57
    boosted, cropped 0.086 / 0.081 0.382 / 0.389 4.44 / 4.82
    stock, full frame 0.120 / 0.128 0.393 / 0.400 3.27 / 3.13
    stock, cropped 0.111 / 0.121 0.389 / 0.396 3.49 / 3.28
    • Cropping lowers measured K in every regime — seeded frames by ~4%, unseeded by ~10% — while intensity rises slightly (e.g. light 243→248): the dropped right-edge columns are dimmer (vignetted), and that intensity nonuniformity was inflating σ/µ beyond true speckle contrast. Unseeded frames benefit more because their K is small enough for the nonuniformity term to matter proportionally more.
    • Intensity congruence is unaffected (boosted/light = 1.037 L / 1.029 R).
    • Net: crop slightly sharpens the seeded/unseeded contrast separation (4.2→4.4 L, 4.6→4.8 R).
    • Crop verified real: every frame's histogram totals 2,201,606 (1720×1280) vs 2,457,606 full-frame.

    Two structural defects this session exposed and fixed (both bench-only discoveries)

    1. Mid-scan rate-LL re-arm was deterministically latching the safety FPGA. The seedless trigger override (LaserPulseSkipDelayUsec=2500) makes each dark frame's return interval 22.5 ms — below the 23.125 ms baseline RATE_LL — so the frame-N restore's re-arm guaranteed a rate_lower_limit_fail latch at the next dark frame of EVERY seedless scan (visible as the trailing safety_se=4 in all previous telemetry; normal scans clear this limit by only ~75 µs). The mid-scan restore now defers the rate-LL re-arm to the teardown restore, which runs after stop_trigger.
    2. The 50 ms fault-clear pulse was marginal on the EE FPGA — it failed twice in a row, and the stale latch starved two scans to 0 frames (TA_shutdown held; fsync also stops ≈2 counts in, mechanism console-side, worth a firmware look). The clear now asserts 300 ms, verifies via STATUS (0x24) read-back, retries 3×, and apply() aborts pre-trigger if the fault won't clear.

    Also found (SDK-side, pre-existing): firmware configure_camera_sensor short-circuits when isConfigured && isPowered, so debug-flag changes (crop/raw) do NOT take effect on already-up cameras — a camera power-cycle before configure is required. Worth documenting on the sensor-fw side.

    Commits: 5e10167 (apply-lock, rebased), 8e0…/latest fix: verify+retry safety-fault clear; defer rate-LL re-arm to teardown, spec updates. Suite: 573 passed, 0 failed. All laser/safety registers verified restored after every run.

  3. boringethan commented on Jul 18, 2026

    @boringethan
    ContributorAuthor

    Per-camera, pipeline-corrected contrast (refines the pooled quick-look numbers above)

    Cameras = mask 0x66 → positions 1, 2, 5, 6 per side (pipeline gains g = 4, 2, 2, 4). Corrections now mirror the real pipeline: pedestal 128 (pedestal_for_fw for FW > 1.5.2), dark correction with variance decomposition (dark.py), and shot-noise subtraction var −= ADC_GAIN·mean·g_cam (shot_noise.py, ADC_GAIN = 0.0815 DN/e⁻). Dark baseline = each scan's 10 seed-off preamble frames — justified because the NRM scan's true terminal dark reads u1 = 128.0 (= pedestal exactly), i.e. seeded CW background ≈ 0 on this bench; 15 s scans contain no usable interior scheduled dark (interval 600), and the BST terminal stop frame caught light (stop-timing artifact — do not use as a dark).

    Crop scans (1720×1280), frames 11..25 seedless vs light, K = σ_corrected/µ_dc:

    scan / side / cam (g) seedless µ_dc / K light µ_dc / K separation
    BST L1(4) 256.6 / 0.0690 246.7 / 0.3261 4.7
    BST L2(2) 239.6 / 0.0895 232.2 / 0.4327 4.8
    BST L5(2) 278.3 / 0.0893 266.0 / 0.4343 4.9
    BST L6(4) 251.2 / 0.0669 244.1 / 0.3280 4.9
    BST R1(4) 174.1 / 0.0645 169.0 / 0.3300 5.1
    BST R2(2) 186.9 / 0.0689 182.9 / 0.4439 6.4
    BST R5(2) 189.4 / 0.0771 183.8 / 0.4420 5.7
    BST R6(4) 175.5 / 0.0662 169.9 / 0.3313 5.0

    (NRM regime table in the analysis script; seedless K per camera matches BST within ~0.01.)

    Observations for the scientists:

    1. Seeded light K splits cleanly by gain family: g=4 cams (1,6) ≈ 0.33; g=2 cams (2,5) ≈ 0.44 — a real optical/speckle-sampling difference the pooled 0.39 was averaging over.
    2. Unseeded K is nearly camera-independent after shot subtraction (0.065–0.090 across all 8 cam-sides) — consistent with a genuine low-coherence contrast floor rather than optics-dependent speckle.
    3. Shot-noise subtraction matters most exactly where it should: seedless frames on g=4 cams drop ~12% (0.079→0.069) vs <1% on bright seeded frames — K²_shot/K² is largest when K is small.
    4. Corrected seedless K is the same in stock and boosted regimes (e.g. L1: 0.0707 vs 0.0690) — K is intensity-independent once shot noise is removed, which is the internal-consistency check on the transfer-function experiment: the boost buys SNR and mean-congruence, not contrast.
    5. Fully-corrected separation: 4.7–6.8× per camera.

    Not applied vs. the full pipeline: dark-interval interpolation (n/a in 15 s scans), calibration C_min/C_max and BFI/BVI conversion, timestamp repair (irrelevant to K).

  4. boringethan commented on Jul 18, 2026

    @boringethan
    ContributorAuthor

    All-8-camera C_min / C_max measurement (Ethan's framing: seedless → C_min, static-phantom light → C_max)

    Per Ethan: the purpose of the seedless frames is to determine C_min, and the light frames — on a static phantom — measure C_max. Full 8-camera coverage per module, boosted seedless regime, crop on, sensor-fw 1.8.2-dev.1, pipeline-corrected (pedestal 128, dark correction from seed-off preamble, shot-noise subtraction with the per-camera gain map). Per-frame K averaged as average-of-ratios (consistent with the C_max convention fix in 8532853).

    Method note — 16-camera scans stall: full 0xFF/0xFF scans stalled the sensor histogram streams (left died at frame 51, right at 89; a rerun at frame 0; uniform across cameras = firmware TX stall, not a camera fault). This is the known parked backpressure limit (ticket #116 / PR #113). All-8 coverage was captured instead as two half-mask scans — 0x66 (cams 1,2,5,6) + 0x99 (cams 0,3,4,7) — same phantom, same session; 4 cams/side streams flawlessly (~3020 chunks, 604/604 frames, RESTORE ALL OK on every run).

    side cam gain C_min (seedless K) C_max (light K) C_max/C_min
    left 0 16 0.0745 0.3271 4.4
    left 1 4 0.0690 0.3261 4.7
    left 2 2 0.0895 0.4327 4.8
    left 3 1 0.0967 0.5189 5.4
    left 4 1 0.1049 0.5233 5.0
    left 5 2 0.0893 0.4343 4.9
    left 6 4 0.0669 0.3280 4.9
    left 7 16 0.0750 0.3325 4.4
    right 0 16 0.0780 0.3362 4.3
    right 1 4 0.0645 0.3300 5.1
    right 2 2 0.0689 0.4439 6.4
    right 3 1 0.1087 0.5171 4.8
    right 4 1 0.0986 0.5240 5.3
    right 5 2 0.0771 0.4420 5.7
    right 6 4 0.0662 0.3313 5.0
    right 7 16 0.0856 0.3452 4.0

    Structure worth the scientists' attention:

    • C_max orders cleanly by gain family: g=1 (cams 3,4) ≈ 0.52 · g=2 (2,5) ≈ 0.43–0.44 · g=4 (1,6) ≈ 0.33 · g=16 (0,7) ≈ 0.33–0.35. The center (low-gain) cameras see the highest speckle contrast; the g=16 edge cams land at the g=4 level, not lower.
    • C_min is much flatter (0.065–0.109) but not featureless: the g=1 cams show the highest C_min (0.097–0.109) — their shot-noise term is smallest, so more residual variance survives the correction; whether the remainder is instrumental or a real coherence floor difference is a question for the team.
    • Separation is 4.0–6.4× on every camera — a usable C_min/C_max span across the whole array.
    • Seedless (C_min) intensity on the g=16 edge cams runs ~97–154 DN — dimmer but still well-resolved above the preamble dark baseline.

    Data: science_out/ scans SEEDLESS_{NRM,BST}_CROP (mask 0x66) and SEEDLESS_{NRM,BST}_CROP_M99 (mask 0x99), with matching telemetry CSVs. Self-lase ≤ ~2 s per scan throughout (watchdog armed, never needed on the half-mask runs).