feat(preset): apply trigger config in SimulatedTxDevice.set_preset - #40

Merged
peterhollender merged 1 commit into
feature/fda_versionfrom
feature/set-preset-applies-trigger-config
Oct 1, 2026
Merged

peterhollender merged 1 commit into
feature/fda_versionfrom
feature/set-preset-applies-trigger-config

Conversation

@openlifu-agent

Copy link
Copy Markdown
Contributor

Companion to the OpenwaterHealth/openlifu-operator-interface#78 smoke-test regression surfaced on OI PR #97: SimulatedTxDevice.set_preset verified the CRC and recorded the loaded selection but did not update _sequence / _pulse with the loaded preset's parameters, so an immediate SimulatedLIFUInterface.start_sonication ran the default 1-train sequence instead of the operator's selected duration. The UI symptom is "ProgressBar never advances; sim reports STOPPED after one tick."

Real firmware doesn't have this problem: on-device set_preset loads the machine_config and applies its registers / profiles, and no separate set_trigger call is needed before start_trigger. The sim now mirrors that model.

Change

SimulatedTxDevice.set_preset now reads from the flash-baked machine_config and applies to the TX device state before returning:

  • pulse_interval_ms (ms → seconds → _sequence["pulse_interval"])
  • pulse_count → _sequence["pulse_count"]
  • pulse_train_interval_s → _sequence["pulse_train_interval"]
  • pulse_train_count_selections[sequence_duration_index] → _sequence["pulse_train_count"]
  • pulse_length_us (μs → seconds → _pulse["duration"])

Also adds a bounds check that raises ValueError when sequence_duration_index is beyond the preset's own pulse_train_count_selections length (previously only the negative-index case raised; an out-of-range positive index silently stored a nonsense selection).

The CRC-mismatch error path is unchanged — no partial state leaks into _sequence / _pulse on a failed CRC check.

Tests

New unit-test/test_simulated_set_preset_applies_trigger.py — 7 tests:

  • set_preset writes pulse_interval / pulse_count / pulse_train_interval / pulse_train_count into _sequence
  • set_preset writes pulse_length_us (converted to seconds) into _pulse["duration"]
  • Applied values visible through get_trigger() — wrapper round-trip
  • Out-of-range sequence_duration_index raises ValueError with "out of range" message
  • Negative sequence_duration_index still raises (regression)
  • CRC mismatch leaves _sequence and _pulse unchanged (no partial state on fault path)
  • Integration: load preset → start_sonication → engine ticks run the selected pulse_train_count, reach trigger_status == STOPPED at train_count == pulse_train_count

Updated _mc() helper in test_simulated_preset_flash.py to include trigger-config fields with a default pulse_train_count_selections=[1, 2, 3, 4, 5] so pre-existing tests calling set_preset(duration_index=0..4) don't need per-site overrides.

Full suite (excluding pre-existing test_tx_device.py collection failure on unrelated HW_ID_DATA_LENGTH import): 114 passed (was 107 pre-change).

Refs OpenwaterHealth/openlifu-operator-interface#78

Self-review

  • Assumption I'm making: the machine_config shape OI builds in preset_schema.to_machine_config is the shape the sim will see. That mapping is: pulse_interval_ms / pulse_count / pulse_train_interval_s / pulse_train_count_selections / pulse_length_us. If the real firmware's machine_config diverges (uses Trigger* prefixed names, say), this needs a translation layer. For now sim and OI are the single source of truth.
  • Alternative considered: have OI call interface.txdevice.set_trigger(...) explicitly after set_preset succeeds, keeping the sim unchanged. Rejected — that leaks "trigger config is a function of preset+duration" knowledge to the host, when the whole point of the SR-003 machine_config handshake is that the firmware (and its sim) is the authoritative source of trigger parameters per loaded preset.
  • New dependency introduced: none.
  • Test coverage: SR-003 fault path (CRC mismatch leaves state untouched); out-of-range duration index raises cleanly; end-to-end start_sonication runs the loaded sequence. Not covered: real-hardware TxDevice.set_preset applying the trigger config via OW_CTRL_SET_PRESET payload (PROJ-04 firmware companion; blocking only for real-hardware verification, not for OI walking skeleton).
  • Breaking change: additive-with-fault-guard. Pre-existing callers of set_preset with a complete machine_config (which OI already builds) just see progress advance correctly now. Callers with an incomplete machine_config (test helpers only, as far as I audited) hit the new ValueError on pulse_train_count_selections bounds — those get updated in the test-helper patch. If any downstream consumer builds machine_configs without pulse_train_count_selections, they'll start getting ValueError on set_preset — none found in grep.
  • Tech-debt filed: none. PROJ-04 firmware extension noted but not debt.

Companion to the OI #78 smoke-test regression. The old
``SimulatedTxDevice.set_preset`` verified the CRC and recorded
the loaded selection but did not update ``_sequence`` /
``_pulse`` with the loaded preset's parameters, so an immediate
``SimulatedLIFUInterface.start_sonication`` ran the default
1-train sequence instead of the operator's selected duration.
That path reaches the UI as "Progress bar never advances; sim
reports STOPPED after one tick."

Real firmware does not have this problem: ``set_preset`` on the
device loads the machine_config and applies its registers /
profiles, and no separate ``set_trigger`` call is needed before
``start_trigger``. The sim now mirrors that model.

## Change

``SimulatedTxDevice.set_preset`` now reads the following fields
from the flash-baked machine_config and applies them to the TX
device state before returning:

* ``pulse_interval_ms`` (ms -> seconds -> ``_sequence["pulse_interval"]``)
* ``pulse_count`` -> ``_sequence["pulse_count"]``
* ``pulse_train_interval_s`` -> ``_sequence["pulse_train_interval"]``
* ``pulse_train_count_selections[sequence_duration_index]`` ->
  ``_sequence["pulse_train_count"]``
* ``pulse_length_us`` (us -> seconds -> ``_pulse["duration"]``)

Also adds a bounds check that raises ``ValueError`` when
``sequence_duration_index`` is beyond the preset's own
``pulse_train_count_selections`` length (previously only the
negative-index case raised; an out-of-range positive index
silently stored a nonsense selection).

The CRC-mismatch error path is unchanged -- no partial state
leaks into ``_sequence`` / ``_pulse`` on a failed CRC check.

## Tests

New: ``unit-test/test_simulated_set_preset_applies_trigger.py``
(7 tests):

* ``set_preset`` writes pulse_interval / pulse_count /
  pulse_train_interval / pulse_train_count into ``_sequence``.
* ``set_preset`` writes pulse_length_us (converted to seconds)
  into ``_pulse["duration"]``.
* The applied values are visible through ``get_trigger()`` --
  wrapper round-trip test.
* Out-of-range ``sequence_duration_index`` raises ``ValueError``
  mentioning "out of range".
* Negative ``sequence_duration_index`` still raises (regression).
* CRC mismatch leaves ``_sequence`` and ``_pulse`` unchanged
  (no partial state on fault path).
* **Integration**: load preset -> start_sonication -> engine ticks
  run the selected ``pulse_train_count``, reach ``trigger_status
  == STOPPED`` at ``train_count == pulse_train_count``.

Updated ``_mc()`` helper in
``unit-test/test_simulated_preset_flash.py`` to include the
required trigger-config fields plus a default
``pulse_train_count_selections=[1, 2, 3, 4, 5]`` so pre-existing
tests that call ``set_preset(duration_index=0..4)`` don't need
per-site overrides.

Full suite (excluding pre-existing ``test_tx_device.py``
collection failure on unrelated ``HW_ID_DATA_LENGTH`` import):
114 passed (was 107 pre-change).

Refs OpenwaterHealth/openlifu-operator-interface#78

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
@peterhollender
peterhollender merged commit 093185f into feature/fda_version Oct 1, 2026
2 checks passed
peterhollender added a commit that referenced this pull request Oct 9, 2026
Companion to the OI #78 smoke-test regression. The old
``SimulatedTxDevice.set_preset`` verified the CRC and recorded
the loaded selection but did not update ``_sequence`` /
``_pulse`` with the loaded preset's parameters, so an immediate
``SimulatedLIFUInterface.start_sonication`` ran the default
1-train sequence instead of the operator's selected duration.
That path reaches the UI as "Progress bar never advances; sim
reports STOPPED after one tick."

Real firmware does not have this problem: ``set_preset`` on the
device loads the machine_config and applies its registers /
profiles, and no separate ``set_trigger`` call is needed before
``start_trigger``. The sim now mirrors that model.

## Change

``SimulatedTxDevice.set_preset`` now reads the following fields
from the flash-baked machine_config and applies them to the TX
device state before returning:

* ``pulse_interval_ms`` (ms -> seconds -> ``_sequence["pulse_interval"]``)
* ``pulse_count`` -> ``_sequence["pulse_count"]``
* ``pulse_train_interval_s`` -> ``_sequence["pulse_train_interval"]``
* ``pulse_train_count_selections[sequence_duration_index]`` ->
  ``_sequence["pulse_train_count"]``
* ``pulse_length_us`` (us -> seconds -> ``_pulse["duration"]``)

Also adds a bounds check that raises ``ValueError`` when
``sequence_duration_index`` is beyond the preset's own
``pulse_train_count_selections`` length (previously only the
negative-index case raised; an out-of-range positive index
silently stored a nonsense selection).

The CRC-mismatch error path is unchanged -- no partial state
leaks into ``_sequence`` / ``_pulse`` on a failed CRC check.

## Tests

New: ``unit-test/test_simulated_set_preset_applies_trigger.py``
(7 tests):

* ``set_preset`` writes pulse_interval / pulse_count /
  pulse_train_interval / pulse_train_count into ``_sequence``.
* ``set_preset`` writes pulse_length_us (converted to seconds)
  into ``_pulse["duration"]``.
* The applied values are visible through ``get_trigger()`` --
  wrapper round-trip test.
* Out-of-range ``sequence_duration_index`` raises ``ValueError``
  mentioning "out of range".
* Negative ``sequence_duration_index`` still raises (regression).
* CRC mismatch leaves ``_sequence`` and ``_pulse`` unchanged
  (no partial state on fault path).
* **Integration**: load preset -> start_sonication -> engine ticks
  run the selected ``pulse_train_count``, reach ``trigger_status
  == STOPPED`` at ``train_count == pulse_train_count``.

Updated ``_mc()`` helper in
``unit-test/test_simulated_preset_flash.py`` to include the
required trigger-config fields plus a default
``pulse_train_count_selections=[1, 2, 3, 4, 5]`` so pre-existing
tests that call ``set_preset(duration_index=0..4)`` don't need
per-site overrides.

Full suite (excluding pre-existing ``test_tx_device.py``
collection failure on unrelated ``HW_ID_DATA_LENGTH`` import):
114 passed (was 107 pre-change).

Refs OpenwaterHealth/openlifu-operator-interface#78

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
Co-authored-by: Peter Hollender <peterhollender@gmail.com>
Signed-off-by: Peter (Agent) <phollender@openwater.cc>
peterhollender added a commit that referenced this pull request Oct 9, 2026
Brings in #38-#40 (simulated preset flash, TrainCount/trigger_status
polling, preset-applied trigger config) on top of the FDA preset
generation work, reconciled to the real OW_PRESET_GET/OW_PRESET_LOAD
contract:

- TxDevice: keep the real get_preset(index) -> dict and
  load_preset(index, regs_crc, duration_index). Drop the
  NotImplementedError get_preset/set_preset placeholders, which
  otherwise silently shadow the real get_preset (later def wins).
- Keep get_trigger train_count / trigger_status (additive, defaulted).
- SimulatedTxDevice owns the (machine_config, regs_crc) flash table;
  get_preset/load_preset match the real shapes; load_preset applies
  the preset's trigger + pulse config (#40 behaviour). Drop the
  set_preset / tuple get_preset / get_machine_config /
  set_preset_flash handshake. Default flash is one loadable preset
  matching the HV default; preset_flash=[] simulates an empty image.
- Single _fw_version for set_version / update_firmware on TX + HV.
- Port test_simulated_preset_flash and the trigger-application tests
  (renamed test_simulated_load_preset_applies_trigger) to load_preset.

Refs SR-002
Refs OpenwaterHealth/openlifu-diathermy-application#16

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
peterhollender added a commit that referenced this pull request Oct 12, 2026
…erge fda_version (#45)

* feat(preset): add simulated preset-integrity handshake endpoints (#38)

* feat(preset): add simulated preset-integrity handshake endpoints

Adds the SDK side of the SR-002 / SR-003 host<->firmware preset-
integrity handshake described in the operator-interface repo's
``preset_pseudo_code.md``.

**Simulated side (implemented):**

* ``SimulatedTxDevice`` now takes an optional ``preset_flash:
  list[tuple[str, int]]`` in its constructor and exposes
  ``get_preset(preset_index) -> (id, settings_crc)``,
  ``set_preset(preset_index, sequence_duration_index, expected_crc)``,
  ``set_preset_flash(flash)``, and ``get_loaded_preset() ->
  (preset_index | None, duration_index | None)`` for test
  round-tripping.

  ``set_preset`` verifies the caller's ``expected_crc`` against the
  flash-baked ``settings_crc``, matching what real firmware will do
  once PROJ-04 lands. Mismatch raises ``ValueError`` -- this is the
  CRC-mismatch fault path the operator UI must annunciate per
  SR-003.

* ``SimulatedLIFUInterface.__init__`` gains a ``preset_flash``
  keyword that is forwarded to the underlying
  ``SimulatedTxDevice``.

**Real side (stubbed):**

* ``TxDevice.get_preset`` / ``TxDevice.set_preset`` are added as
  documented ``NotImplementedError`` stubs pointing at PROJ-04 for
  firmware work. Their docstrings direct testers to
  ``SimulatedLIFUInterface`` in the meantime.

**Tests (new, ``unit-test/test_simulated_preset_flash.py``):**

* Happy-path: ``get_preset`` returns flash entries; ``set_preset``
  records selection on matching CRC; ``get_loaded_preset``
  round-trips.
* Rejection: ``set_preset`` on CRC mismatch / out-of-range index /
  negative duration index; ``get_preset`` out-of-range.
* Flash reseed clears loaded selection.
* ``SimulatedLIFUInterface.__init__`` forwards ``preset_flash`` to
  its txdevice.
* ``TxDevice.get_preset`` / ``set_preset`` raise
  ``NotImplementedError`` mentioning PROJ-04.

All 14 new tests pass. Full ``unit-test/`` run (excluding the
pre-existing ``test_tx_device.py`` collection failure that is
unrelated to this change): 92 passed.

No firmware changes in this PR. The corresponding operator-
interface work that consumes this API lands in
OpenwaterHealth/openlifu-operator-interface as a follow-up (SR-002
/ SR-003 UI wiring; part of PROJ-03 close-out).

Signed-off-by: Peter Hollender <peterhollender@gmail.com>

* feat(sim): add version mutators for firmware-compat simulation

SimulatedTxDevice.set_version(version, module=0) and SimulatedHVController.set_version(version) let a debug UI or test flip the reported firmware version so the operator connector's check_firmware_compat can be driven down the incompatible-firmware path without shipping a separate `firmware_compatible` toggle. Single string per device today; can extend to per-module later.

3 new tests, 17 total in test_simulated_preset_flash.py.

Signed-off-by: Peter Hollender <peterhollender@gmail.com>

* refactor(sim): preset_flash holds machine_config dicts; implement update_firmware

Two related SDK-side changes that support the SR-003 preset-integrity
handshake per the operator-interface repo's ``preset_pseudo_code.md``:

**1. preset_flash schema change (breaking)**

The simulated flash-baked preset table now holds full
``machine_config`` dicts, not just ``(id, settings_crc)`` tuples.
Each entry is ``(machine_config_dict, machine_config_crc)`` where:

* ``machine_config_dict`` mirrors what the real firmware image
  would carry after PROJ-04 codegen -- specifically, includes an
  ``id`` and a ``settings_crc`` field the ``get_preset`` method
  reads.
* ``machine_config_crc`` is what the host sends to ``set_preset``
  at runtime -- the CRC of the full machine_config, not the CRC
  of the underlying preset dict.

New method: ``SimulatedTxDevice.get_machine_config(preset_index)``
returns ``(machine_config_dict, machine_config_crc)`` for
debugging + tests to inspect the flash-side state.

``set_preset``'s CRC-mismatch error message now names
``machine_config_crc`` explicitly so callers understand which
CRC was checked (the machine_config's, not the preset's).

**2. Simulator's update_firmware is implemented**

``SimulatedTxDevice.update_firmware`` and
``SimulatedHVController.update_firmware`` now flip the reported
firmware version to a ``target_version`` kwarg (default:
``"sim-1.0.7"``). Real DFU is out of scope for the simulator; the
purpose here is to let an app-side firmware-update flow (e.g. the
operator-interface's "Firmware update required" modal ->
``connector.update_firmware()`` -> re-probe) run end-to-end in
sim mode.

The signatures mirror the shape of the real
``TxDevice.update_firmware`` / HV equivalent so app-side code can
call the same method against either backend.

Tests: 22 total (was 17). New coverage:

* ``machine_config_dict`` -> ``(id, settings_crc)`` extraction
* ``get_machine_config`` round-trip
* ``set_preset`` CRC-mismatch error message names
  ``machine_config_crc``
* ``update_firmware`` on both TX + HV flips ``get_version`` output

No firmware changes in this PR.

Signed-off-by: Peter Hollender <peterhollender@gmail.com>

---------

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
Co-authored-by: Peter Hollender <peterhollender@gmail.com>
Signed-off-by: Peter (Agent) <phollender@openwater.cc>

* feat(trigger): add TrainCount + trigger_status for polled progress (#39)

Lets the operator interface poll sonication progress against
``TxDevice.get_trigger()`` instead of subscribing to the async
``signal_data_received`` STATUS-frame OWSignal. Companion to
openlifu-operator-interface#78 (PROJ-08 Task 1: Running-state
ProgressBar + firmware-driven Running->Finished transition).

The walking-skeleton OI deliberately dropped async telemetry /
status-frame subscriptions in openlifu-operator-interface#92;
reintroducing one for sonication progress would undo that work.
A polled model fits the sync-probe idiom the OI state machine
already uses for connect state and firmware compat.

## Changes

### Simulated TxDevice

* ``SimulatedTxDevice.__init__`` adds ``self._train_curr = 0``
  so a polled host can read progress without subscribing to the
  OWSignal. Written by the engine on each tick.
* ``SimulatedTxDevice.get_trigger_json()`` adds ``"TrainCount":
  self._train_curr`` to its returned dict.
* ``SimulatedTxDevice.get_trigger()`` is new -- mirrors the shape
  of the real ``LIFUTXDevice.TxDevice.get_trigger()`` so OI code
  calling ``interface.txdevice.get_trigger()`` gets the same keys
  against sim and real hardware. Adds two walking-skeleton
  polling fields: ``train_count`` (current running counter) and
  ``trigger_status`` (``"RUNNING"`` / ``"STOPPED"``).

### Sim engine

* ``_SimulatedRunEngine.start()`` resets ``self._tx._train_curr =
  0`` so a polled host sees progress starting from 0 on every
  fresh sonication (previously the counter carried over from a
  prior run if the host polled before the first tick).
* ``_SimulatedRunEngine._on_train_tick`` mirrors
  ``self._train_curr`` back to ``self._tx._train_curr`` after
  each increment -- the field the polled host reads.
* ``_SimulatedRunEngine._on_train_tick`` natural-completion
  branch now calls ``self._tx.stop_trigger()`` (previously only
  an explicit ``stop_sonication`` flipped ``_trigger_running``
  to False, so a polled host saw ``TriggerStatus=RUNNING`` after
  the sim finished naturally). The emitted STATUS frame already
  carried ``status=STOPPED`` so callers that subscribed to the
  OWSignal weren't affected by the bug; polled callers were.

### Real TxDevice

* ``LIFUTXDevice.get_trigger()`` extends its return dict with
  the same two fields:
  * ``"train_count": trigger_json.get("TrainCount", 0)``
  * ``"trigger_status": trigger_json.get("TriggerStatus",
    "UNKNOWN")``
  Both use ``.get(default)`` so pre-PROJ-04 firmware (which
  doesn't yet populate ``TrainCount``) returns 0 cleanly. The
  firmware-side ``OW_CTRL_GET_SWTRIG`` response-payload
  extension to actually populate the field is tracked
  separately as a PROJ-04 companion on
  openlifu-transmitter-fw -- not blocking the OI walking
  skeleton since OI runs against the sim pre-PROJ-04.

## Tests

New: ``unit-test/test_simulated_trigger_progress.py`` -- 7 tests
exercising the sim-side changes end-to-end:

* ``get_trigger_json`` / ``get_trigger`` initial state (both
  ``TrainCount`` / ``train_count`` default to 0, ``TriggerStatus``
  / ``trigger_status`` default to ``STOPPED``).
* ``get_trigger`` snake-case shape check (every expected key
  present).
* ``train_count`` advances monotonically during a short synthetic
  sequence (5 pulse trains * 20 ms), reaches ``pulse_train_count``
  on natural completion.
* Natural completion flips ``trigger_status`` to ``STOPPED`` --
  regression test for the pre-2026-10-01 bug where
  ``_trigger_running`` wasn't flipped on natural end.
* Operator-initiated ``stop_sonication`` midway: ``trigger_status``
  flips to ``STOPPED``, ``train_count`` retained at partial
  progress (not reset to 0, not advanced to total).
* Restart resets ``train_count`` to 0 at engine ``start()``.

Full suite (excluding pre-existing ``test_tx_device.py`` collection
failure on an unrelated ``HW_ID_DATA_LENGTH`` import): 107 passed.

Refs OpenwaterHealth/openlifu-operator-interface#78

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
Co-authored-by: Peter Hollender <peterhollender@gmail.com>
Signed-off-by: Peter (Agent) <phollender@openwater.cc>

* feat(preset): apply trigger config in SimulatedTxDevice.set_preset (#40)

Companion to the OI #78 smoke-test regression. The old
``SimulatedTxDevice.set_preset`` verified the CRC and recorded
the loaded selection but did not update ``_sequence`` /
``_pulse`` with the loaded preset's parameters, so an immediate
``SimulatedLIFUInterface.start_sonication`` ran the default
1-train sequence instead of the operator's selected duration.
That path reaches the UI as "Progress bar never advances; sim
reports STOPPED after one tick."

Real firmware does not have this problem: ``set_preset`` on the
device loads the machine_config and applies its registers /
profiles, and no separate ``set_trigger`` call is needed before
``start_trigger``. The sim now mirrors that model.

## Change

``SimulatedTxDevice.set_preset`` now reads the following fields
from the flash-baked machine_config and applies them to the TX
device state before returning:

* ``pulse_interval_ms`` (ms -> seconds -> ``_sequence["pulse_interval"]``)
* ``pulse_count`` -> ``_sequence["pulse_count"]``
* ``pulse_train_interval_s`` -> ``_sequence["pulse_train_interval"]``
* ``pulse_train_count_selections[sequence_duration_index]`` ->
  ``_sequence["pulse_train_count"]``
* ``pulse_length_us`` (us -> seconds -> ``_pulse["duration"]``)

Also adds a bounds check that raises ``ValueError`` when
``sequence_duration_index`` is beyond the preset's own
``pulse_train_count_selections`` length (previously only the
negative-index case raised; an out-of-range positive index
silently stored a nonsense selection).

The CRC-mismatch error path is unchanged -- no partial state
leaks into ``_sequence`` / ``_pulse`` on a failed CRC check.

## Tests

New: ``unit-test/test_simulated_set_preset_applies_trigger.py``
(7 tests):

* ``set_preset`` writes pulse_interval / pulse_count /
  pulse_train_interval / pulse_train_count into ``_sequence``.
* ``set_preset`` writes pulse_length_us (converted to seconds)
  into ``_pulse["duration"]``.
* The applied values are visible through ``get_trigger()`` --
  wrapper round-trip test.
* Out-of-range ``sequence_duration_index`` raises ``ValueError``
  mentioning "out of range".
* Negative ``sequence_duration_index`` still raises (regression).
* CRC mismatch leaves ``_sequence`` and ``_pulse`` unchanged
  (no partial state on fault path).
* **Integration**: load preset -> start_sonication -> engine ticks
  run the selected ``pulse_train_count``, reach ``trigger_status
  == STOPPED`` at ``train_count == pulse_train_count``.

Updated ``_mc()`` helper in
``unit-test/test_simulated_preset_flash.py`` to include the
required trigger-config fields plus a default
``pulse_train_count_selections=[1, 2, 3, 4, 5]`` so pre-existing
tests that call ``set_preset(duration_index=0..4)`` don't need
per-site overrides.

Full suite (excluding pre-existing ``test_tx_device.py``
collection failure on unrelated ``HW_ID_DATA_LENGTH`` import):
114 passed (was 107 pre-change).

Refs OpenwaterHealth/openlifu-operator-interface#78

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
Co-authored-by: Peter Hollender <peterhollender@gmail.com>
Signed-off-by: Peter (Agent) <phollender@openwater.cc>

* refactor(io): split DeviceInterface (FDA) from LIFUInterface (RUO)

DeviceInterface is now the FDA-facing base class: connection,
monitoring, PID lock, load_preset (TX load + HV select with
frequency-specific sensitivity), start/stop sonication, status, close.
LIFUInterface extends it with the research-only endpoints
(set_solution, check_solution, set_module_invert, voltage tables,
sequence duty/duration) and the RUO constructor options
(ext_power_supply, module_invert, voltage_table_selection) via a
_create_devices hook. Existing LIFUInterface callers are unchanged.

Sensitivity lookup is a single sensitivity_at_frequency() in
LIFUUserConfig (np.interp per module, clamped, mean across modules);
TxDevice and SimulatedTxDevice both delegate to it.

HVController.select_preset sends <I (CRC only) or <Iff (CRC,
sensitivity, frequency_hz); a half-specified call raises ValueError.
Removed the unused HVController.load_preset alias.

Simulator mirrors the split: SimulatedDeviceInterface exposes no RUO
methods; SimulatedLIFUInterface adds them. Simulated preset paths raise
LIFUDeviceError (OW_BAD_CRC / OW_HV_PRESET_CRC) and LIFUSolutionError
like the hardware path.

Refs SR-002
Refs OpenwaterHealth/openlifu-diathermy-application#16

Signed-off-by: Peter Hollender <peterhollender@gmail.com>

* refactor(io): split FDA TxDevice/HVController from RUO subclasses (#47)

* refactor(io): split FDA TxDevice/HVController from RUO subclasses

DeviceInterface (the FDA entry point) handed out TxDevice and
HVController objects that still carried every RUO command (register
writes, set_voltage, set_trigger, DFU, config writes, async mode).
Restricting the interface alone left them one attribute away.

- TxDevice / HVController are now the FDA components. They expose only
  identity, telemetry, trigger read/start/stop and stored-preset
  methods, and carry a SUPPORTED_COMMANDS whitelist
  (TX_FDA_COMMANDS / HV_FDA_COMMANDS) so any other opcode raises
  ValueError before it reaches the UART.
- LIFUTxDevice / LIFUHVController subclass them and add the RUO
  methods; shared RUO globals (start/stop, send_async, echo, LED,
  reset, DFU, write_config) move to LIFUComponentMixin.
- Only LIFUInterface creates the LIFU classes (_create_devices hook).
- Async mode is RUO-only: DeviceInterface drops run_async,
  start_monitoring/stop_monitoring and the async_mode calls in
  start/stop_sonication; LIFUInterface keeps them.
- The simulator mirrors the split (SimulatedLIFUTxDevice,
  SimulatedLIFUHVController; monitoring and async_mode on
  SimulatedLIFUInterface).
- RUO unit tests now use the LIFU classes; new tests cover the class
  surfaces, the opcode whitelist and which classes each interface
  builds. test_tx_device.py imports HW_ID_DATA_LENGTH from LIFUConfig
  so it collects again (three stale set_trigger tests remain, #44).

Refs SR-002
Closes #43
Refs #44

Signed-off-by: Peter Hollender <peterhollender@gmail.com>

* test(tx): update set_trigger train-interval tests to back-to-back contract

test_23b/23c/23d asserted the 2.0.14 behaviour (auto-filling a zero
train interval and bumping it past the pulse period for firmware
<= 2.0.3). Since 8b3eaf2 set_trigger sends 0 for back-to-back trains,
and firmware <= 2.0.3 no longer needs support.

Replace them with tests of the current contract: zero or zero-gap
intervals are sent as 0; longer intervals pass through; a short
interval is sent as 0 for a single train and raises ValueError (with
nothing sent) for multiple trains.

Refs SR-002
Closes #44

Signed-off-by: Peter Hollender <peterhollender@gmail.com>

---------

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
Co-authored-by: Peter Hollender <peterhollender@gmail.com>
Signed-off-by: Peter (Agent) <phollender@openwater.cc>

---------

Signed-off-by: Peter Hollender <peterhollender@gmail.com>
Signed-off-by: Peter (Agent) <phollender@openwater.cc>
Co-authored-by: Peter Hollender <peterhollender@gmail.com>
peterhollender pushed a commit that referenced this pull request Oct 12, 2026
#45 was squash-merged into feature/fda-preset-generation, so the
#38-#40 commits it carried are not ancestors of this branch, and a
merge into feature/fda_version produced 17 spurious conflict hunks
even though the content is already here.

feature/fda_version was first reset to the DCO-signed copies of
#38-#40 (identical trees, same parent). This merge uses the "ours"
strategy to add that tip as a parent without changing any file, so
the follow-up PR into feature/fda_version is conflict-free.

Refs SR-002
Refs #45

Signed-off-by: Peter (Agent) <phollender@openwater.cc>