feat(preset): apply trigger config in SimulatedTxDevice.set_preset - #40
Merged
peterhollender merged 1 commit intoOct 1, 2026
Merged
peterhollender merged 1 commit into
peterhollender merged 1 commit into
Conversation
Companion to the OI #78 smoke-test regression. The old ``SimulatedTxDevice.set_preset`` verified the CRC and recorded the loaded selection but did not update ``_sequence`` / ``_pulse`` with the loaded preset's parameters, so an immediate ``SimulatedLIFUInterface.start_sonication`` ran the default 1-train sequence instead of the operator's selected duration. That path reaches the UI as "Progress bar never advances; sim reports STOPPED after one tick." Real firmware does not have this problem: ``set_preset`` on the device loads the machine_config and applies its registers / profiles, and no separate ``set_trigger`` call is needed before ``start_trigger``. The sim now mirrors that model. ## Change ``SimulatedTxDevice.set_preset`` now reads the following fields from the flash-baked machine_config and applies them to the TX device state before returning: * ``pulse_interval_ms`` (ms -> seconds -> ``_sequence["pulse_interval"]``) * ``pulse_count`` -> ``_sequence["pulse_count"]`` * ``pulse_train_interval_s`` -> ``_sequence["pulse_train_interval"]`` * ``pulse_train_count_selections[sequence_duration_index]`` -> ``_sequence["pulse_train_count"]`` * ``pulse_length_us`` (us -> seconds -> ``_pulse["duration"]``) Also adds a bounds check that raises ``ValueError`` when ``sequence_duration_index`` is beyond the preset's own ``pulse_train_count_selections`` length (previously only the negative-index case raised; an out-of-range positive index silently stored a nonsense selection). The CRC-mismatch error path is unchanged -- no partial state leaks into ``_sequence`` / ``_pulse`` on a failed CRC check. ## Tests New: ``unit-test/test_simulated_set_preset_applies_trigger.py`` (7 tests): * ``set_preset`` writes pulse_interval / pulse_count / pulse_train_interval / pulse_train_count into ``_sequence``. * ``set_preset`` writes pulse_length_us (converted to seconds) into ``_pulse["duration"]``. * The applied values are visible through ``get_trigger()`` -- wrapper round-trip test. * Out-of-range ``sequence_duration_index`` raises ``ValueError`` mentioning "out of range". * Negative ``sequence_duration_index`` still raises (regression). * CRC mismatch leaves ``_sequence`` and ``_pulse`` unchanged (no partial state on fault path). * **Integration**: load preset -> start_sonication -> engine ticks run the selected ``pulse_train_count``, reach ``trigger_status == STOPPED`` at ``train_count == pulse_train_count``. Updated ``_mc()`` helper in ``unit-test/test_simulated_preset_flash.py`` to include the required trigger-config fields plus a default ``pulse_train_count_selections=[1, 2, 3, 4, 5]`` so pre-existing tests that call ``set_preset(duration_index=0..4)`` don't need per-site overrides. Full suite (excluding pre-existing ``test_tx_device.py`` collection failure on unrelated ``HW_ID_DATA_LENGTH`` import): 114 passed (was 107 pre-change). Refs OpenwaterHealth/openlifu-operator-interface#78 Signed-off-by: Peter Hollender <peterhollender@gmail.com>
peterhollender
approved these changes
Oct 1, 2026
peterhollender
added a commit
that referenced
this pull request
Oct 9, 2026
Companion to the OI #78 smoke-test regression. The old ``SimulatedTxDevice.set_preset`` verified the CRC and recorded the loaded selection but did not update ``_sequence`` / ``_pulse`` with the loaded preset's parameters, so an immediate ``SimulatedLIFUInterface.start_sonication`` ran the default 1-train sequence instead of the operator's selected duration. That path reaches the UI as "Progress bar never advances; sim reports STOPPED after one tick." Real firmware does not have this problem: ``set_preset`` on the device loads the machine_config and applies its registers / profiles, and no separate ``set_trigger`` call is needed before ``start_trigger``. The sim now mirrors that model. ## Change ``SimulatedTxDevice.set_preset`` now reads the following fields from the flash-baked machine_config and applies them to the TX device state before returning: * ``pulse_interval_ms`` (ms -> seconds -> ``_sequence["pulse_interval"]``) * ``pulse_count`` -> ``_sequence["pulse_count"]`` * ``pulse_train_interval_s`` -> ``_sequence["pulse_train_interval"]`` * ``pulse_train_count_selections[sequence_duration_index]`` -> ``_sequence["pulse_train_count"]`` * ``pulse_length_us`` (us -> seconds -> ``_pulse["duration"]``) Also adds a bounds check that raises ``ValueError`` when ``sequence_duration_index`` is beyond the preset's own ``pulse_train_count_selections`` length (previously only the negative-index case raised; an out-of-range positive index silently stored a nonsense selection). The CRC-mismatch error path is unchanged -- no partial state leaks into ``_sequence`` / ``_pulse`` on a failed CRC check. ## Tests New: ``unit-test/test_simulated_set_preset_applies_trigger.py`` (7 tests): * ``set_preset`` writes pulse_interval / pulse_count / pulse_train_interval / pulse_train_count into ``_sequence``. * ``set_preset`` writes pulse_length_us (converted to seconds) into ``_pulse["duration"]``. * The applied values are visible through ``get_trigger()`` -- wrapper round-trip test. * Out-of-range ``sequence_duration_index`` raises ``ValueError`` mentioning "out of range". * Negative ``sequence_duration_index`` still raises (regression). * CRC mismatch leaves ``_sequence`` and ``_pulse`` unchanged (no partial state on fault path). * **Integration**: load preset -> start_sonication -> engine ticks run the selected ``pulse_train_count``, reach ``trigger_status == STOPPED`` at ``train_count == pulse_train_count``. Updated ``_mc()`` helper in ``unit-test/test_simulated_preset_flash.py`` to include the required trigger-config fields plus a default ``pulse_train_count_selections=[1, 2, 3, 4, 5]`` so pre-existing tests that call ``set_preset(duration_index=0..4)`` don't need per-site overrides. Full suite (excluding pre-existing ``test_tx_device.py`` collection failure on unrelated ``HW_ID_DATA_LENGTH`` import): 114 passed (was 107 pre-change). Refs OpenwaterHealth/openlifu-operator-interface#78 Signed-off-by: Peter Hollender <peterhollender@gmail.com> Co-authored-by: Peter Hollender <peterhollender@gmail.com> Signed-off-by: Peter (Agent) <phollender@openwater.cc>
peterhollender
added a commit
that referenced
this pull request
Oct 9, 2026
Brings in #38-#40 (simulated preset flash, TrainCount/trigger_status polling, preset-applied trigger config) on top of the FDA preset generation work, reconciled to the real OW_PRESET_GET/OW_PRESET_LOAD contract: - TxDevice: keep the real get_preset(index) -> dict and load_preset(index, regs_crc, duration_index). Drop the NotImplementedError get_preset/set_preset placeholders, which otherwise silently shadow the real get_preset (later def wins). - Keep get_trigger train_count / trigger_status (additive, defaulted). - SimulatedTxDevice owns the (machine_config, regs_crc) flash table; get_preset/load_preset match the real shapes; load_preset applies the preset's trigger + pulse config (#40 behaviour). Drop the set_preset / tuple get_preset / get_machine_config / set_preset_flash handshake. Default flash is one loadable preset matching the HV default; preset_flash=[] simulates an empty image. - Single _fw_version for set_version / update_firmware on TX + HV. - Port test_simulated_preset_flash and the trigger-application tests (renamed test_simulated_load_preset_applies_trigger) to load_preset. Refs SR-002 Refs OpenwaterHealth/openlifu-diathermy-application#16 Signed-off-by: Peter Hollender <peterhollender@gmail.com>
peterhollender
added a commit
that referenced
this pull request
Oct 12, 2026
…erge fda_version (#45) * feat(preset): add simulated preset-integrity handshake endpoints (#38) * feat(preset): add simulated preset-integrity handshake endpoints Adds the SDK side of the SR-002 / SR-003 host<->firmware preset- integrity handshake described in the operator-interface repo's ``preset_pseudo_code.md``. **Simulated side (implemented):** * ``SimulatedTxDevice`` now takes an optional ``preset_flash: list[tuple[str, int]]`` in its constructor and exposes ``get_preset(preset_index) -> (id, settings_crc)``, ``set_preset(preset_index, sequence_duration_index, expected_crc)``, ``set_preset_flash(flash)``, and ``get_loaded_preset() -> (preset_index | None, duration_index | None)`` for test round-tripping. ``set_preset`` verifies the caller's ``expected_crc`` against the flash-baked ``settings_crc``, matching what real firmware will do once PROJ-04 lands. Mismatch raises ``ValueError`` -- this is the CRC-mismatch fault path the operator UI must annunciate per SR-003. * ``SimulatedLIFUInterface.__init__`` gains a ``preset_flash`` keyword that is forwarded to the underlying ``SimulatedTxDevice``. **Real side (stubbed):** * ``TxDevice.get_preset`` / ``TxDevice.set_preset`` are added as documented ``NotImplementedError`` stubs pointing at PROJ-04 for firmware work. Their docstrings direct testers to ``SimulatedLIFUInterface`` in the meantime. **Tests (new, ``unit-test/test_simulated_preset_flash.py``):** * Happy-path: ``get_preset`` returns flash entries; ``set_preset`` records selection on matching CRC; ``get_loaded_preset`` round-trips. * Rejection: ``set_preset`` on CRC mismatch / out-of-range index / negative duration index; ``get_preset`` out-of-range. * Flash reseed clears loaded selection. * ``SimulatedLIFUInterface.__init__`` forwards ``preset_flash`` to its txdevice. * ``TxDevice.get_preset`` / ``set_preset`` raise ``NotImplementedError`` mentioning PROJ-04. All 14 new tests pass. Full ``unit-test/`` run (excluding the pre-existing ``test_tx_device.py`` collection failure that is unrelated to this change): 92 passed. No firmware changes in this PR. The corresponding operator- interface work that consumes this API lands in OpenwaterHealth/openlifu-operator-interface as a follow-up (SR-002 / SR-003 UI wiring; part of PROJ-03 close-out). Signed-off-by: Peter Hollender <peterhollender@gmail.com> * feat(sim): add version mutators for firmware-compat simulation SimulatedTxDevice.set_version(version, module=0) and SimulatedHVController.set_version(version) let a debug UI or test flip the reported firmware version so the operator connector's check_firmware_compat can be driven down the incompatible-firmware path without shipping a separate `firmware_compatible` toggle. Single string per device today; can extend to per-module later. 3 new tests, 17 total in test_simulated_preset_flash.py. Signed-off-by: Peter Hollender <peterhollender@gmail.com> * refactor(sim): preset_flash holds machine_config dicts; implement update_firmware Two related SDK-side changes that support the SR-003 preset-integrity handshake per the operator-interface repo's ``preset_pseudo_code.md``: **1. preset_flash schema change (breaking)** The simulated flash-baked preset table now holds full ``machine_config`` dicts, not just ``(id, settings_crc)`` tuples. Each entry is ``(machine_config_dict, machine_config_crc)`` where: * ``machine_config_dict`` mirrors what the real firmware image would carry after PROJ-04 codegen -- specifically, includes an ``id`` and a ``settings_crc`` field the ``get_preset`` method reads. * ``machine_config_crc`` is what the host sends to ``set_preset`` at runtime -- the CRC of the full machine_config, not the CRC of the underlying preset dict. New method: ``SimulatedTxDevice.get_machine_config(preset_index)`` returns ``(machine_config_dict, machine_config_crc)`` for debugging + tests to inspect the flash-side state. ``set_preset``'s CRC-mismatch error message now names ``machine_config_crc`` explicitly so callers understand which CRC was checked (the machine_config's, not the preset's). **2. Simulator's update_firmware is implemented** ``SimulatedTxDevice.update_firmware`` and ``SimulatedHVController.update_firmware`` now flip the reported firmware version to a ``target_version`` kwarg (default: ``"sim-1.0.7"``). Real DFU is out of scope for the simulator; the purpose here is to let an app-side firmware-update flow (e.g. the operator-interface's "Firmware update required" modal -> ``connector.update_firmware()`` -> re-probe) run end-to-end in sim mode. The signatures mirror the shape of the real ``TxDevice.update_firmware`` / HV equivalent so app-side code can call the same method against either backend. Tests: 22 total (was 17). New coverage: * ``machine_config_dict`` -> ``(id, settings_crc)`` extraction * ``get_machine_config`` round-trip * ``set_preset`` CRC-mismatch error message names ``machine_config_crc`` * ``update_firmware`` on both TX + HV flips ``get_version`` output No firmware changes in this PR. Signed-off-by: Peter Hollender <peterhollender@gmail.com> --------- Signed-off-by: Peter Hollender <peterhollender@gmail.com> Co-authored-by: Peter Hollender <peterhollender@gmail.com> Signed-off-by: Peter (Agent) <phollender@openwater.cc> * feat(trigger): add TrainCount + trigger_status for polled progress (#39) Lets the operator interface poll sonication progress against ``TxDevice.get_trigger()`` instead of subscribing to the async ``signal_data_received`` STATUS-frame OWSignal. Companion to openlifu-operator-interface#78 (PROJ-08 Task 1: Running-state ProgressBar + firmware-driven Running->Finished transition). The walking-skeleton OI deliberately dropped async telemetry / status-frame subscriptions in openlifu-operator-interface#92; reintroducing one for sonication progress would undo that work. A polled model fits the sync-probe idiom the OI state machine already uses for connect state and firmware compat. ## Changes ### Simulated TxDevice * ``SimulatedTxDevice.__init__`` adds ``self._train_curr = 0`` so a polled host can read progress without subscribing to the OWSignal. Written by the engine on each tick. * ``SimulatedTxDevice.get_trigger_json()`` adds ``"TrainCount": self._train_curr`` to its returned dict. * ``SimulatedTxDevice.get_trigger()`` is new -- mirrors the shape of the real ``LIFUTXDevice.TxDevice.get_trigger()`` so OI code calling ``interface.txdevice.get_trigger()`` gets the same keys against sim and real hardware. Adds two walking-skeleton polling fields: ``train_count`` (current running counter) and ``trigger_status`` (``"RUNNING"`` / ``"STOPPED"``). ### Sim engine * ``_SimulatedRunEngine.start()`` resets ``self._tx._train_curr = 0`` so a polled host sees progress starting from 0 on every fresh sonication (previously the counter carried over from a prior run if the host polled before the first tick). * ``_SimulatedRunEngine._on_train_tick`` mirrors ``self._train_curr`` back to ``self._tx._train_curr`` after each increment -- the field the polled host reads. * ``_SimulatedRunEngine._on_train_tick`` natural-completion branch now calls ``self._tx.stop_trigger()`` (previously only an explicit ``stop_sonication`` flipped ``_trigger_running`` to False, so a polled host saw ``TriggerStatus=RUNNING`` after the sim finished naturally). The emitted STATUS frame already carried ``status=STOPPED`` so callers that subscribed to the OWSignal weren't affected by the bug; polled callers were. ### Real TxDevice * ``LIFUTXDevice.get_trigger()`` extends its return dict with the same two fields: * ``"train_count": trigger_json.get("TrainCount", 0)`` * ``"trigger_status": trigger_json.get("TriggerStatus", "UNKNOWN")`` Both use ``.get(default)`` so pre-PROJ-04 firmware (which doesn't yet populate ``TrainCount``) returns 0 cleanly. The firmware-side ``OW_CTRL_GET_SWTRIG`` response-payload extension to actually populate the field is tracked separately as a PROJ-04 companion on openlifu-transmitter-fw -- not blocking the OI walking skeleton since OI runs against the sim pre-PROJ-04. ## Tests New: ``unit-test/test_simulated_trigger_progress.py`` -- 7 tests exercising the sim-side changes end-to-end: * ``get_trigger_json`` / ``get_trigger`` initial state (both ``TrainCount`` / ``train_count`` default to 0, ``TriggerStatus`` / ``trigger_status`` default to ``STOPPED``). * ``get_trigger`` snake-case shape check (every expected key present). * ``train_count`` advances monotonically during a short synthetic sequence (5 pulse trains * 20 ms), reaches ``pulse_train_count`` on natural completion. * Natural completion flips ``trigger_status`` to ``STOPPED`` -- regression test for the pre-2026-10-01 bug where ``_trigger_running`` wasn't flipped on natural end. * Operator-initiated ``stop_sonication`` midway: ``trigger_status`` flips to ``STOPPED``, ``train_count`` retained at partial progress (not reset to 0, not advanced to total). * Restart resets ``train_count`` to 0 at engine ``start()``. Full suite (excluding pre-existing ``test_tx_device.py`` collection failure on an unrelated ``HW_ID_DATA_LENGTH`` import): 107 passed. Refs OpenwaterHealth/openlifu-operator-interface#78 Signed-off-by: Peter Hollender <peterhollender@gmail.com> Co-authored-by: Peter Hollender <peterhollender@gmail.com> Signed-off-by: Peter (Agent) <phollender@openwater.cc> * feat(preset): apply trigger config in SimulatedTxDevice.set_preset (#40) Companion to the OI #78 smoke-test regression. The old ``SimulatedTxDevice.set_preset`` verified the CRC and recorded the loaded selection but did not update ``_sequence`` / ``_pulse`` with the loaded preset's parameters, so an immediate ``SimulatedLIFUInterface.start_sonication`` ran the default 1-train sequence instead of the operator's selected duration. That path reaches the UI as "Progress bar never advances; sim reports STOPPED after one tick." Real firmware does not have this problem: ``set_preset`` on the device loads the machine_config and applies its registers / profiles, and no separate ``set_trigger`` call is needed before ``start_trigger``. The sim now mirrors that model. ## Change ``SimulatedTxDevice.set_preset`` now reads the following fields from the flash-baked machine_config and applies them to the TX device state before returning: * ``pulse_interval_ms`` (ms -> seconds -> ``_sequence["pulse_interval"]``) * ``pulse_count`` -> ``_sequence["pulse_count"]`` * ``pulse_train_interval_s`` -> ``_sequence["pulse_train_interval"]`` * ``pulse_train_count_selections[sequence_duration_index]`` -> ``_sequence["pulse_train_count"]`` * ``pulse_length_us`` (us -> seconds -> ``_pulse["duration"]``) Also adds a bounds check that raises ``ValueError`` when ``sequence_duration_index`` is beyond the preset's own ``pulse_train_count_selections`` length (previously only the negative-index case raised; an out-of-range positive index silently stored a nonsense selection). The CRC-mismatch error path is unchanged -- no partial state leaks into ``_sequence`` / ``_pulse`` on a failed CRC check. ## Tests New: ``unit-test/test_simulated_set_preset_applies_trigger.py`` (7 tests): * ``set_preset`` writes pulse_interval / pulse_count / pulse_train_interval / pulse_train_count into ``_sequence``. * ``set_preset`` writes pulse_length_us (converted to seconds) into ``_pulse["duration"]``. * The applied values are visible through ``get_trigger()`` -- wrapper round-trip test. * Out-of-range ``sequence_duration_index`` raises ``ValueError`` mentioning "out of range". * Negative ``sequence_duration_index`` still raises (regression). * CRC mismatch leaves ``_sequence`` and ``_pulse`` unchanged (no partial state on fault path). * **Integration**: load preset -> start_sonication -> engine ticks run the selected ``pulse_train_count``, reach ``trigger_status == STOPPED`` at ``train_count == pulse_train_count``. Updated ``_mc()`` helper in ``unit-test/test_simulated_preset_flash.py`` to include the required trigger-config fields plus a default ``pulse_train_count_selections=[1, 2, 3, 4, 5]`` so pre-existing tests that call ``set_preset(duration_index=0..4)`` don't need per-site overrides. Full suite (excluding pre-existing ``test_tx_device.py`` collection failure on unrelated ``HW_ID_DATA_LENGTH`` import): 114 passed (was 107 pre-change). Refs OpenwaterHealth/openlifu-operator-interface#78 Signed-off-by: Peter Hollender <peterhollender@gmail.com> Co-authored-by: Peter Hollender <peterhollender@gmail.com> Signed-off-by: Peter (Agent) <phollender@openwater.cc> * refactor(io): split DeviceInterface (FDA) from LIFUInterface (RUO) DeviceInterface is now the FDA-facing base class: connection, monitoring, PID lock, load_preset (TX load + HV select with frequency-specific sensitivity), start/stop sonication, status, close. LIFUInterface extends it with the research-only endpoints (set_solution, check_solution, set_module_invert, voltage tables, sequence duty/duration) and the RUO constructor options (ext_power_supply, module_invert, voltage_table_selection) via a _create_devices hook. Existing LIFUInterface callers are unchanged. Sensitivity lookup is a single sensitivity_at_frequency() in LIFUUserConfig (np.interp per module, clamped, mean across modules); TxDevice and SimulatedTxDevice both delegate to it. HVController.select_preset sends <I (CRC only) or <Iff (CRC, sensitivity, frequency_hz); a half-specified call raises ValueError. Removed the unused HVController.load_preset alias. Simulator mirrors the split: SimulatedDeviceInterface exposes no RUO methods; SimulatedLIFUInterface adds them. Simulated preset paths raise LIFUDeviceError (OW_BAD_CRC / OW_HV_PRESET_CRC) and LIFUSolutionError like the hardware path. Refs SR-002 Refs OpenwaterHealth/openlifu-diathermy-application#16 Signed-off-by: Peter Hollender <peterhollender@gmail.com> * refactor(io): split FDA TxDevice/HVController from RUO subclasses (#47) * refactor(io): split FDA TxDevice/HVController from RUO subclasses DeviceInterface (the FDA entry point) handed out TxDevice and HVController objects that still carried every RUO command (register writes, set_voltage, set_trigger, DFU, config writes, async mode). Restricting the interface alone left them one attribute away. - TxDevice / HVController are now the FDA components. They expose only identity, telemetry, trigger read/start/stop and stored-preset methods, and carry a SUPPORTED_COMMANDS whitelist (TX_FDA_COMMANDS / HV_FDA_COMMANDS) so any other opcode raises ValueError before it reaches the UART. - LIFUTxDevice / LIFUHVController subclass them and add the RUO methods; shared RUO globals (start/stop, send_async, echo, LED, reset, DFU, write_config) move to LIFUComponentMixin. - Only LIFUInterface creates the LIFU classes (_create_devices hook). - Async mode is RUO-only: DeviceInterface drops run_async, start_monitoring/stop_monitoring and the async_mode calls in start/stop_sonication; LIFUInterface keeps them. - The simulator mirrors the split (SimulatedLIFUTxDevice, SimulatedLIFUHVController; monitoring and async_mode on SimulatedLIFUInterface). - RUO unit tests now use the LIFU classes; new tests cover the class surfaces, the opcode whitelist and which classes each interface builds. test_tx_device.py imports HW_ID_DATA_LENGTH from LIFUConfig so it collects again (three stale set_trigger tests remain, #44). Refs SR-002 Closes #43 Refs #44 Signed-off-by: Peter Hollender <peterhollender@gmail.com> * test(tx): update set_trigger train-interval tests to back-to-back contract test_23b/23c/23d asserted the 2.0.14 behaviour (auto-filling a zero train interval and bumping it past the pulse period for firmware <= 2.0.3). Since 8b3eaf2 set_trigger sends 0 for back-to-back trains, and firmware <= 2.0.3 no longer needs support. Replace them with tests of the current contract: zero or zero-gap intervals are sent as 0; longer intervals pass through; a short interval is sent as 0 for a single train and raises ValueError (with nothing sent) for multiple trains. Refs SR-002 Closes #44 Signed-off-by: Peter Hollender <peterhollender@gmail.com> --------- Signed-off-by: Peter Hollender <peterhollender@gmail.com> Co-authored-by: Peter Hollender <peterhollender@gmail.com> Signed-off-by: Peter (Agent) <phollender@openwater.cc> --------- Signed-off-by: Peter Hollender <peterhollender@gmail.com> Signed-off-by: Peter (Agent) <phollender@openwater.cc> Co-authored-by: Peter Hollender <peterhollender@gmail.com>
peterhollender
pushed a commit
that referenced
this pull request
Oct 12, 2026
#45 was squash-merged into feature/fda-preset-generation, so the #38-#40 commits it carried are not ancestors of this branch, and a merge into feature/fda_version produced 17 spurious conflict hunks even though the content is already here. feature/fda_version was first reset to the DCO-signed copies of #38-#40 (identical trees, same parent). This merge uses the "ours" strategy to add that tip as a parent without changing any file, so the follow-up PR into feature/fda_version is conflict-free. Refs SR-002 Refs #45 Signed-off-by: Peter (Agent) <phollender@openwater.cc>
2 tasks
This file contains hidden or bidirectional Unicode text that may be interpreted or compiled differently than what appears below. To review, open the file in an editor that reveals hidden Unicode characters.
Learn more about bidirectional Unicode characters
Sign up for free
to join this conversation on GitHub.
Already have an account?
Sign in to comment
Add this suggestion to a batch that can be applied as a single commit.This suggestion is invalid because no changes were made to the code.Suggestions cannot be applied while the pull request is closed.Suggestions cannot be applied while viewing a subset of changes.Only one suggestion per line can be applied in a batch.Add this suggestion to a batch that can be applied as a single commit.Applying suggestions on deleted lines is not supported.You must change the existing code in this line in order to create a valid suggestion.Outdated suggestions cannot be applied.This suggestion has been applied or marked resolved.Suggestions cannot be applied from pending reviews.Suggestions cannot be applied on multi-line comments.Suggestions cannot be applied while the pull request is queued to merge.Suggestion cannot be applied right now. Please check back later.
Companion to the OpenwaterHealth/openlifu-operator-interface#78 smoke-test regression surfaced on OI PR #97:
SimulatedTxDevice.set_presetverified the CRC and recorded the loaded selection but did not update_sequence/_pulsewith the loaded preset's parameters, so an immediateSimulatedLIFUInterface.start_sonicationran the default 1-train sequence instead of the operator's selected duration. The UI symptom is "ProgressBar never advances; sim reports STOPPED after one tick."Real firmware doesn't have this problem: on-device
set_presetloads the machine_config and applies its registers / profiles, and no separateset_triggercall is needed beforestart_trigger. The sim now mirrors that model.Change
SimulatedTxDevice.set_presetnow reads from the flash-baked machine_config and applies to the TX device state before returning:pulse_interval_ms(ms → seconds →_sequence["pulse_interval"])pulse_count→_sequence["pulse_count"]pulse_train_interval_s→_sequence["pulse_train_interval"]pulse_train_count_selections[sequence_duration_index]→_sequence["pulse_train_count"]pulse_length_us(μs → seconds →_pulse["duration"])Also adds a bounds check that raises
ValueErrorwhensequence_duration_indexis beyond the preset's ownpulse_train_count_selectionslength (previously only the negative-index case raised; an out-of-range positive index silently stored a nonsense selection).The CRC-mismatch error path is unchanged — no partial state leaks into
_sequence/_pulseon a failed CRC check.Tests
New
unit-test/test_simulated_set_preset_applies_trigger.py— 7 tests:set_presetwrites pulse_interval / pulse_count / pulse_train_interval / pulse_train_count into_sequenceset_presetwrites pulse_length_us (converted to seconds) into_pulse["duration"]get_trigger()— wrapper round-tripsequence_duration_indexraisesValueErrorwith "out of range" messagesequence_duration_indexstill raises (regression)_sequenceand_pulseunchanged (no partial state on fault path)start_sonication→ engine ticks run the selectedpulse_train_count, reachtrigger_status == STOPPEDattrain_count == pulse_train_countUpdated
_mc()helper intest_simulated_preset_flash.pyto include trigger-config fields with a defaultpulse_train_count_selections=[1, 2, 3, 4, 5]so pre-existing tests callingset_preset(duration_index=0..4)don't need per-site overrides.Full suite (excluding pre-existing
test_tx_device.pycollection failure on unrelatedHW_ID_DATA_LENGTHimport): 114 passed (was 107 pre-change).Refs OpenwaterHealth/openlifu-operator-interface#78
Self-review
preset_schema.to_machine_configis the shape the sim will see. That mapping is:pulse_interval_ms/pulse_count/pulse_train_interval_s/pulse_train_count_selections/pulse_length_us. If the real firmware's machine_config diverges (usesTrigger*prefixed names, say), this needs a translation layer. For now sim and OI are the single source of truth.interface.txdevice.set_trigger(...)explicitly afterset_presetsucceeds, keeping the sim unchanged. Rejected — that leaks "trigger config is a function of preset+duration" knowledge to the host, when the whole point of the SR-003 machine_config handshake is that the firmware (and its sim) is the authoritative source of trigger parameters per loaded preset.TxDevice.set_presetapplying the trigger config viaOW_CTRL_SET_PRESETpayload (PROJ-04 firmware companion; blocking only for real-hardware verification, not for OI walking skeleton).set_presetwith a complete machine_config (which OI already builds) just see progress advance correctly now. Callers with an incomplete machine_config (test helpers only, as far as I audited) hit the newValueErroronpulse_train_count_selectionsbounds — those get updated in the test-helper patch. If any downstream consumer builds machine_configs withoutpulse_train_count_selections, they'll start gettingValueErroronset_preset— none found in grep.